Meteorological grade instrument for
measurement of biologically effective ultraviolet
radiation(UV-B) outdoors and underwater (5m max depth)

APPLICATIONS
- measurement of biologically effective ultraviolet
radiation from sunlight and from artificial sources with
wavelengths above 280 nm
- plant, marine, climatological ozone or other biological
impact studies that require continuous field measurements
over long periods of time or to detect trends
- correction of spectroradiometric measurement errors cause
by the variations of irradiance during the scan
- sunburning time measurements for media and research
organizations that provide public information about the
effects of UV radiation
CONFIGURATION
Model 501 is a complete plug-and-play system including the
detector for outdoor use, computerized controller/data logger
and connecting cables. It is easy to install and use. 
Model 501A is a standalone detector that can be connected to
any data logger. Recommended for sites with existing data
collection equipment or requiring data loggers designed for
outdoor use. Detector cable is provided.
SPECIFICATIONS
Detectors of Model 501 and 501A
- Spectral range:
- 290-320 nm, close to Erythema Action Spectrum

- Measurement range:
- 0 to 10 MED/Hr (Minimal Erythemal Dose per Hour)
- Angular response:
- within 5% from ideal cosine for incident angles
- Response time:
- 1 second (0.1 second on demand)
- Temperature error:
- RAF:
- >1 (Radiation Amplification Factor calculated for 30
deg solar zenith angle and 0.27 cm ozone column)
- Oper. temperature:
- -40 to +50 degC ambient
- Environment:
- outdoors or underwater (max. depth 15 feet or 5 m)
Recorder of Model 501 only
- Temperature range:
- 0 to +50C
- Environment:
- indoors, no water condensation
FEATURES
Model 501
- Output:
- digital (UV signal and sensor temperature)
- Data acquisition:
- dedicated computerized data logger with liquid crystal
display, keypad, serial communication port and parallel
printer port; accommodates 2 detectors simultaneously;
optional analog output
- Dose period:
- selectable from 1 minute to 1 hour
- Internal memory:
- stores up to six month of 1 hour dose integrals and sensor
temperature from two detectors in a nonvolatile memory
- Data integrity:
- nonvolatile memory; internal rechargeable power backup for
the recording circuitry; automatic temperature compensation
corrects data if ambient temperature exceeds specifications
or AC power is lost
- Communication:
- compatible with any computer and modem equipped with an
RS232 serial port; menu organized remote setup and data
retrieval; user selectable baud rate from 300 to 9600 bps
- Power source:
- universal 90-250 VAC, 50-60 Hz or 12 VDC
Model 501A
- Output:
- analog (UV signal and sensor temperature); compatible with
most data loggers
- Power source:
- 11 to 15 VDC/1A max. (5mA without temperature
stabilization)
PHYSICAL CHARACTERISTICS
Detectors of Model 501 and 501A
- Weight:
- 2 lbs (0.9 kg) without cable
- Dimensions:
- 5.5" H x 5.9" Diam. (14 x 15 cm)
- Cable:
- with hermetic connector at detector end; 50 ft standard
length, up to 200 ft. can be ordered; approx. weight 0.1 lbs
per ft
- Mounting:
- 3 leveling legs locked in a quick release base; circular
level mounted on the detector
Recorder of Model 501
- Weight:
- 4 lbs (1.8 kg)
- Dimensions:
- 6.3" W x 3.1" H x 9.5" D (16x8x24 cm)
Back to UVB Radiation